36
Setting up and running the Instrument
9. Adjust the Dispensing Speed
Please note, the tubing cartridge speci cations are set
using the highest speed (S-1)
1. Press the PRIME and EMPTY keys simultaneously
and turn on the instrument.
2. Press SELECT to cycle through the three pump-
speed choices as they appear in the LED Display.
The choices are:
S-1 = High speed (This is the normal setting).
S-2 = Medium speed
S-3 = Low speed
3. When the speed you want to use shows in the
display, turn off the instrument, and on again.