Setting up and running the Instrument
a. Use the UP or DOWN arrows to select the
duration of the pause. The default pause duration
is set to .01 milliseconds.
b. Press the SELECT button to save your setting.
Pump Count (PC): The number on the display represents
how many revolutions the pump has made.
a. Press the SELECT button to clear the pump
count.
b. Press the SELECT button again, when “sure” is
displayed.
Note: If the count exceeds 9999, the count is displayed in
exponential form, such as “10E3” for 10,000 revolutions or
“12E4” for 123,456 revolutions.
Timed Prime (PT): The prime duration is set in real time
seconds from 1-9999.
a. Select the time (for example, 20 for 20 seconds)
using the UP or DOWN arrows
b. Press the SELECT button to save the time.
When the prime button is pressed under normal operation
(and the timed prime is not zero), the system will count
down the time for priming (for example, 20 seconds) rather
than requiring you to hold down the prime button.
Back Step (BS): The back step volume is set in 100
nanosecond units. For example, 0.1 – 20.0µl are available
for back stepping the pump motor after dispense and
before pausing the pump.
a. Use the UP or DOWN arrows to select the
duration of the back step. The default back step is
set to 0.0µl.
b. Press the Select button to save your setting.