46
Troubleshooting
Table 8. Stacker Troubleshooting: Issues and Solutions
Error Issue Solution
E101 Plate stage on the
WellMate is not at
home position.
Turn off WellMate and Stacker
unit. Make sure the plate stage on
the WellMate is all the way to the
right in its home position.
E103 Sensor is reporting that
the right plate stage is not
at the starting lift position.
Make sure that the right plate
stage is in the correct position.
Turn the stacker unit off, move the
plate shuttle to access the right
plate stage, and make sure the
breakaway washer and screw are
set correctly.
E104 Sensor is reporting that
the left plate stage is not
at the starting lift position.
Make sure that the left plate
stage is in the correct position.
Turn the stacker unit off, move
the plate shuttle to access the left
plate stage, and make sure the
breakaway washer and screw are
set correctly.
E105 A plate is on the right
stage while the stage is
lifting up.
Remove the plate from the right
stage.
E106 A plate is on the left stage
while the stage is lifting
up.
Remove the plate from the left
stage.