Manualofuseandmaintenance
EN
65
DANGER:
Donotactivatethedeviceifthehandpieceisfaultyorbroken.Immediatelyreplacethehandpiece.
DANGER:
Iftheperistalticpumpisfaulty,usetheotherchannel.
DANGER: Inserts with water irrigation.
Checktheirrigationreleasebeforeandduringuse.Makesuretheliquidcomesoutoftheinsert’s
point.
Donotusethedeviceifthephysiologicalsolutionisnotreleasedorthepumpisfaulty.
DANGER: Injury to people.
Makesurethecablesdonothinderthefreepassageofpeople.
DANGER: Injury to people.
Whentheperistalticpumpapisopen(Pic.12-Ref.C)donotactivatethePiezosurgeryMedical
footswitch.Themovingpartscouldinjuretheoperator.
WARNING:
Donotusepointedorsharpobjectsonthelcdscreen.Thiscoulddamagethetouchscreenorthe
screenitself.Alightpressureofthenger,eveninalatexglove,allowstheoperatortochoosethe
desiredsettings.
WARNING: Contraindication.
Do not use to treat metal or ceramic prostheses.The ultrasound vibrations could cause the de-
cementationoftheprostheses.
DANGER:
Beforesurgery,makesuretheinsertiscorrectlyxedtothehandpiece.
DANGER:
Checkthattheinsertxedtothehandpieceisthesameastheoneselectedonthedisplay,otherwise
anoptimalperformancemaynotbeobtained.
DANGER:
ThePiezosurgeryMedicaldeviceisandeviceusedinbonesurgery.Howeveraprolongedcontact
and/or excessive force of the inserts on soft tissues should be avoided because these could
causeburnsand/ornoncutlesions.Useparticularcautionwhenusingsharpinserts.Aprolonged
mechanicaluseofsharpinsertscouldcutthroughthesofttissues.Whenusinginproximityofsoft
tissues/nerves(forexampletheperineviumoftheperipheralnervoussystemortheduramaterof
thecentralnervoussystem)itishighlyrecommendedtonishthecutwithabluntdiamondinsertin
ordertominimizethepotentialriskofdamagingthesetissues.