Manualofuseandmaintenance
EN
93
12 TECHNICAL DATA AND ELECTROMAGNETIC
COMPATIBILITY
12.1 Technical data
DeviceconformingtoEECDir.93/42: ClassII.
ClassicationasprovidedforinEN60601-1: ClassI
TypeB
IP20(device)
IPX8(footswitch)
Intermittentfunctiondevice 60”ON30”OFFwithirrigation
30”ON120”OFFwithoutirrigation
Powersupplytension: 100-240Vac50/60Hz.
Max.powerabsorbed: 200VA
Fuses: Type:5X20mmT2AL,250V
Workfrequency: Automaticscanningfrom24KHzto36KHz
Averagepowerlevelsappliedtothehandpiece: 4powervariationsarepossible:
CORTICAL
CANCELLOUSMEDIUM
CANCELLOUSLOW
DELICATEANATOMY
Notallpowervariationscanbeusedforeveryinsert:
thevariationswhichcannotbeselectedarenot
viewed.
Everypowervariationhas7levels.
Thechoiceoflevelisallowedwithinthepreset
intervalestablishedbyMectron,inrelationtothe
insertandpowervariationselected.
Rangeoftheperistalticpump: Thereisascaleof5rangelevels,from7to65ml/min
approximately.
Thechoiceoflevelisallowedwithinthepreset
intervalestablishedbyMectron,inrelationtothe
insertandthetypeofcutselected.
APCcircuitprotections: -Nohandpiece
-Cordwiredisconnected
-Insertnotlockedproperlyorbroken
-Groundingprotectionactivated
Malfunctiondetectionsystem: Thedisplayonthefrontindicatesananomaly
detectedandthecorrectiveactiontotake(seech.10
“Descriptionofshields,alarmsandsolutions”)
Operatingconditions: From+10°Cto+40°C.
Relativehumidityfrom30%to75%.
Transportandstoringconditions: From-10°Cto+70°C.
Relativehumidityfrom10%to90%.
AirpressureP:from500hPato1060hPa
Weightandsize: 11.7Kg
L-l-h410X380X310mm.