5.2.7 Choice of language
Aftertheentiresystemhasbeenchecked,thefollowingdisplayappears:(Pic.11).
Thersttimethedeviceisswitchedon,thepresetlanguageistheEnglishlanguage.Ifadifferent
languageisrequired:
- Pressthe“LANGUAGE”button(Pic.11-Ref.A).
- Selecttheagcorrespondingtothedesiredlanguage.
- Press“RETURN”:theselectedlanguagewillthenbethepresetlanguagewhichwillremainthe
sameuntilanewselectionismade.
NOTE:Thelanguagemaybemodiedonlywhenthethereisnohandpieceinsertedinthedevice.
Pic.11
A
5.2.8 Setting up the irrigation circuit
A. OPERATIONS TO CARRY OUT IN STERILE CONDITIONS:
DANGER:
ThePiezosurgeryMedicalIrrigationKitisguaranteedforasingleuse.Segregateanddisposeofthe
kitaccordingtothelawsinforceregulatinghospitalwaste.
DANGER:
BeforeusingthePiezosurgeryMedicalIrrigationKit,checktheintegrityofthesterilepackingand
carefullyinspecttheproducttoexcludeanyeventualdamageorfaults.Donotusetheirrigationkitif
thepackinghasbeentamperedwithordamaged.
Theirrigationkitlosesitssterileconditionifthepackingisbroken or damaged. If the packing is
damaged,disposeofthekit.Donotattempttosterilizeorreusethekit.