Manualofuseandmaintenance
DANGER: Contraindications.
CheckthatthePiezosurgeryMedicalhandpiecehasbeencleanedandsterilizedbeforeusing.Ifit
isn’t,donotuse.
DANGER:
ThePiezosurgeryMedicalhandpiece(neworserviced)isnotsterilewhenitissuppliedandmust
thereforebecarefullycleanedandsterilized(withtheautoclave)beforeusing.
DANGER:
AftersterilizingthePiezosurgeryMedicalhandpieceintheautoclave,waituntilithascooleddown
beforeusing.
WARNING:
Whenthesterilizationcycleiscompleteandbeforeconnectingthehandpiecetothedevicebody,
makesuretheelectricalcontactsofthehandpiece-cordconnectorareperfectlydry.Ifnecessary,dry
thecontactsbyblowingcompressedairformedicaluse.
DANGER:
After using, each handpiece-cord must be thoroughly cleaned and sterilized before using again,
followingtheinstructionscontainedinthe “CleaningandSterilizingInstructions”manualsupplied
withthePiezosurgeryMedicaldevice.
DANGER:
Alwayscheck,beforeandafterusing,thatthehandpieceanditscordarenotdamaged.Ifdamageis
detected,donotusethehandpiece-cord.