10 DESCRIPTION OF SHIELDS, ALARMS AND SOLUTIONS
Thedeviceisprovidedwithadiagnosticscircuitwhichshowstheactivationoftheshieldsandalarms
onthedisplay.Insomecasesitisnecessarytocontacttechnicalassistance,makingreferenceto
errorcode“MED-XXX”displayed.
Threetypesoferrorcanbeidentied:
1. Blocking:
Alarmswhich for safetyreasonsinterruptallthedevice’soperationsinhibitingtheuseofthe
singlechannelortheentiredevicepreventinganyharmtothepatient.
2. Non-blocking:
Alarmswhich,forsafetyreasons,interruptallthedevice’soperations,momentarilyinhibitingthe
useofthesinglechannelortheentiredevice.Conrmationofacknowledgmentofthealarmis
neededinordertocontinue.
3. Information messages:
Alarmswhichindicatethepresenceofananomaly,butdonotblockthedevicethatcontinuesto
functioneitherinnormalconditionsorinreducedfunctionmode.
NOTE:
Incaseanonblockingerroroccursonthechannelinuse,afterconrmingthealarmreadingitis
necessarytopressthe“START”buttonagaintocontinue.
Message Description Action needed
MED - 001
Replace the handpiece
Malfunctionofthesyntoniccircuit
orfaultyhandpiece.
Replacethehandpiece.
MED - 002
Replace the handpiece
Replacethehandpiece.
MED - 004
Please check that the handpiece
is used properly. If the problem
persists, replace it
Pleasecheckthatthehandpiece
isusedproperly.Iftheproblem
persists,replaceit.
MED - 005
Tighten the insert tip or replace
it
-Insertnotcorrectlyfastenedto
thehandpiece.
-Insertworn,brokenordeformed.
-Inrarecasesthepresenceof
humidityintheelectricalcontacts
ofthecordmayoccur
Tightentheinserttiporreplaceit.
Drytheelectricalcontactsby
blowingcompressedair.
MED - 006
Internal problem. “SHUTDOWN”
the device then start it again
Offrangepowersupplyshield
enabledsignal.
“SHUTDOWN”thedevicethen
startitagain
MED - 016
Channel out of use
Failedcommunicationwiththe
channelmodule
Usethechannelthatworks.
Contacttechnicalassistance.
MED - 017
Make sure that the pump and
tube are correctly installed
Errorinpumpfunctionsignal. -Makesurethatthepumpand
tubearecorrectlyinstalled.
-Makesurethereisnothing
impedingthepump’srotation
MED - 021
Channel out of use
Checksumerrorsignal,rmware
modulealteration.
Usethechannelthatworks.
Contacttechnicalassistance.