Manualofuseandmaintenance
EN
89
Message Description Action needed
MED - 022
Internal problem: please contact
assistance
Checksumerrorsignal,rmware
modulealteration.
-Thedeviceisfunctioning
properly:
contactassoonaspossible
-Thedeviceisnotfunctioning
properly:contacttechnical
assistanceimmediately
MED - 028
External power supply problem.
Check the settings
Anomalyinexternalpowersupply. Checkthatthesettingsarecorrect
andcontinueoperating.
MED - 029
Channel out of use
Moduleinterfaceprogram
executionerrorsignal.
Contacttechnicalassistance
immediately.
MED - 030
IInternal problem: please
contact assistance
Internalprocessexecutionerror
signal.
-Thedeviceisfunctioning
properly:
contactassoonaspossible
-Thedeviceisnotfunctioning
properly:contacttechnical
assistanceimmediately
MED - 032
Lack of internal communication:
please contact assistance
Failedinternalcommunication. -Thedeviceisfunctioning
properly:
contactassoonaspossible
-Thedeviceisnotfunctioning
properly:contacttechnical
assistanceimmediately
MED - 037
Fill the irrigation tubing line, by
pressing the “PUMP” button
Filltheirrigationtubingline,by
pressingthe“PUMP”button
PressthePUMPbuttontollthe
irrigationtube.
MED - 038
“PUMP” button not active. Fill
the irrigation tubing line by
pressing the footswitch
“PUMP”buttonnotactive. Filltheirrigationtubebykeeping
thefootswitchpressed.
MED - 040
Current handpiece
unserviceable. Replace the
handpiece
Currenthandpieceunserviceable. -Replacethehandpiece.
-Requestsoftwareupdatefrom
technicalassistance.
MED - 041
This insert tip cannot be
selected for the handpiece being
used
Thisinserttipcannotbeselected
forthehandpiecebeingused.
ChangetypeofHandpieceor
changeinsert.
MED - 043
Default language - English
Languageselectionnotcorrect:
Englishisthedefaultlanguage.
Tryselectingthelanguage
requiredbyusingthe“Language”
button.Iftheproblempersists,
contacttechnicalassistance.
MED - 044
The data need to be
synchronised. Press “OK” as
soon as possible. The process
lasts a few seconds
Internalproblemofdata
synchronization.
Finishthecurrentoperationand
thenstartsynchronizationassoon
aspossiblebypressingOK.
MED - 046
INFO not available
INFOnotavailableinthelanguage
chosenfortheselectedinsert.
Pressthe“OK”keytocancelthe
messagefromthescreen.