Manualofuseandmaintenance
EN
91
11 PROBLEM-SOLVING
Ifthedevicedoesnotseemtobeworkingproperly,readtheinstructionsagainandthencheckthe
followingtable.
Problem Possible cause Solution
Thedevicedoesnotswitchon
afterpositioningtheswitchon
“ON”.Thegreenledlightisoff
(Pic.7-Ref.E).
Thepowersupplycableterminalis
notinsertedproperlyinthesocket
onthebackofthedevice.
Checkthatthepowersupplycable
isrmlyconnected.
Thepowersupplycableisfaulty. Checkthatthepowersupply
socketisworking.
Replacethepowersupplycable.
Thefusesareoutoforder. Replacethefuses(seechapter7
“Replacingthefuses”).
Thedeviceisswitchedonbut
doesn’twork.Thedisplaydoesnot
showanymessages.
Thefootswitchjackisnotcorrectly
insertedinthesocket.
Insertthefootswitchjackcorrectly
initssocket.
Thefootswitchdoesn’twork. ContactMectronauthorized
technicalassistance.
Whenworkingahissisheard
comingfromthePiezosurgery
Medicalhandpiece.
Theinsertisnotcorrectlylocked
onthehandpiece.
Unscrewtheinsertandscrewback
oncorrectly.
Thedeviceworkscorrectlybutthe
pumpisn’trunningproperly.
Incorrectpositioningofthetubein
theperistalticpump.
Checkthatthetubeisinserted
correctlyintheperistalticpump
(seeparagraph5.2.8).
Thepumpisrunningcorrectlybut
whenitstopsliquidseepsoutof
thehandpiece.
Theperistalticpumpapisnot
properlyshut.
Checkthattheperistalticpump
apisperfectlyshut(see
paragraph5.2.8).
Whenworkingnoliquidcomesout
oftheinsert.
TheIRRIGATIONfunctionisset
zeroed.
Increasethequantityofliquid
todeliverusingthe“+”keyof
theIRRIGATIONfunction(see
paragraph5.3.2Phase3).
Thesolutionbagisempty. Replacethebagwithafullone.
Theairinletorthetubeclampare
notopen.
Opentheairinletandthetube
clamp.
Theirrigationtubingandthepump
tubesarenotcorrectlytted.
Checkthetubeconnections.
Theinsertisobstructed Cleartheuidspassageofthe
insert.
Thehandpieceisobstructed. ContactMectronauthorized
technicalassistance.
Thetubeoftheperistalticpumpis
nottherightsize.
OnlyuseoriginalMectronspares.