11 PROBLEM-SOLVING
Ifthedevicedoesnotseemtobeworkingproperly,readtheinstructionsagainandthencheckthe
followingtable.
Problem Possible cause Solution
Thedevicedoesnotswitchon
afterpositioningtheswitchon
“ON”.Thegreenledlightisoff
(Pic.7-Ref.E).
Thepowersupplycableterminalis
notinsertedproperlyinthesocket
onthebackofthedevice.
Checkthatthepowersupplycable
isrmlyconnected.
Thepowersupplycableisfaulty. Checkthatthepowersupply
socketisworking.
Replacethepowersupplycable.
Thefusesareoutoforder. Replacethefuses(seechapter7
“Replacingthefuses”).
Thedeviceisswitchedonbut
doesn’twork.Thedisplaydoesnot
showanymessages.
Thefootswitchjackisnotcorrectly
insertedinthesocket.
Insertthefootswitchjackcorrectly
initssocket.
Thefootswitchdoesn’twork. ContactMectronauthorized
technicalassistance.
Whenworkingahissisheard
comingfromthePiezosurgery
Medicalhandpiece.
Theinsertisnotcorrectlylocked
onthehandpiece.
Unscrewtheinsertandscrewback
oncorrectly.
Thedeviceworkscorrectlybutthe
pumpisn’trunningproperly.
Incorrectpositioningofthetubein
theperistalticpump.
Checkthatthetubeisinserted
correctlyintheperistalticpump
(seeparagraph5.2.8).
Thepumpisrunningcorrectlybut
whenitstopsliquidseepsoutof
thehandpiece.
Theperistalticpumpapisnot
properlyshut.
Checkthattheperistalticpump
apisperfectlyshut(see
paragraph5.2.8).
Whenworkingnoliquidcomesout
oftheinsert.
TheIRRIGATIONfunctionisset
zeroed.
Increasethequantityofliquid
todeliverusingthe“+”keyof
theIRRIGATIONfunction(see
paragraph5.3.2Phase3).
Thesolutionbagisempty. Replacethebagwithafullone.
Theairinletorthetubeclampare
notopen.
Opentheairinletandthetube
clamp.
Theirrigationtubingandthepump
tubesarenotcorrectlytted.
Checkthetubeconnections.
Theinsertisobstructed Cleartheuidspassageofthe
insert.
Thehandpieceisobstructed. ContactMectronauthorized
technicalassistance.
Thetubeoftheperistalticpumpis
nottherightsize.
OnlyuseoriginalMectronspares.