Recommended separation distances between portable and mobile RF communications equipment and
the Piezosurgery Medical Device
ThePiezosurgeryMedicaldeviceisdesignedtoworkinanelectromagneticenvironmentinwhichRFradiated
disturbancesarekeptundercontrol.ThepurchaseroruserofthePiezosurgeryMedicaldevicecancontribute
topreventelectromagneticinterferencebyensuringaminimumdistancebetweentheRFmobileandcordless
communicationdevice(transmitters)andthePiezosurgeryMedicaldevice,asrecommendedbelow,inrelation
tothemaximumoutputpoweroftheradio-communicationdevices.
Ratedmaximumoutput
poweroftransmitter“W”
Separationdistanceaccordingtofrequencyoftransmitter“m”
150kHzto80MHz
d = 1,2 √P
80MHzto800MHz
d = 1,2 √P
800MHzto2,5GHz
d = 2,3 √P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
Fortransmitterswithamaximumnominaloutput powernotindicatedintheabovetable,therecommended
separationdistancedinmeters(m)canbecalculatedusingtheequationapplicabletothefrequencyofthe
transmitter,wherePisthemaximumnominaloutputpowerofthetransmitterinWatts(W) accordingtothe
manufacturerofthetransmitter.
Note:
(1)At80MHzand800MHzthehighestfrequencyintervalisapplied.
(2) These guidelines might not apply to all situations. Electromagnetic propagation is inuenced by the
absorptionandreectionofstructures,objectsandpeople.