Digital Color Doppler Ultrasound System
182
Basic User Manual
Electromagnetic emissions
The EQUIPME NT is intended for use in the electromagnetic environm ent specified below. The
customer or the user of the EQUIPMENT should assure that it is us ed in such an environment.
Emissions test
Compliance
Electromagnetic environment
-
guidance
The EQUIPMENT uses RF energy only for its internal
RF emissions
CISPR 11
Group 1
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The EQUIPMENT is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low -voltage
power s upply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
Class A
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
Complies
IEC 61000-3-3