Digital Color Doppler Ultrasound System
181
Basic User Manual
Appendix A
EMC Guidance and Manufacturer's Declaration
This product complies with the EMC standard IEC 60601-1-2:2007.
Note:
The system should not be used adjacent to or stacked with other equipm ent. If the
adjacent or tacked use is nec essary, the system should be observed to verify normal
operation in the configuration in which it will be used.
The system needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided below.
Preventing conducted RF immunity. Due t o technological limitations, the conducted RF
immunity level are limited to 1Vrms level, conducted RF interference above 1Vrms may
cause wrong diagnosis and measurem ents. We suggested that you position the system
further from sources of conducted RF noise.
Portable and mobile RF communications equipment can affects this system.