84 Medit T-series User Guide
5.3 Electromagnetic Compatibility Information
Electro-Magnetic Emissions
This MD-ID0400 is intended for use in the electromagnetic environment as specified below. The
customer or the user of the MD-ID0400 should ensure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions CISPR
11
Group 1
The MD-ID0400 uses RF energy only for its internal functions.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions CISPR
11
Class A
The EUT is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/ Flicker
emissions
Complies
Warning: This MD-ID0400 is intended for use by healthcare professionals only. This equipment/ system may cause
radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures,
such as re-orienting or relocating the MD-ID0400 or shielding the location.
Electro-Magnetic Immunity
Guidance 1
This MD-ID0400 is intended for use in the electromagnetic environment as specified below. The
customer or the user of the MD-ID0400 system should ensure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact ± 2
kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 8 kV contact ± 2
kV, ± 4 kV, ± 8 kV,
± 15 kV air
Floors should be of wood, concrete or
ceramic tiles. If floors are covered with a
synthetic material, the relative humidity
should be at least 30% is
recommended.