21 Medit T-series User Guide
Medit Scan for Labs: Instruction
3.1 Calibration: Table Top and Intraoral Scanners
Calibration is recommended for proper scanning and performance of the device.
Please calibrate the device if:
The quality of scan data has decreased as compared to the previous scans.
The external conditions such as the temperature of device has changed during use.
If it is already past the configured calibration period.
To set the calibration period, go to Menu > Settings > Calibration Period (Days) option.
The calibration panel is a very sensitive component
Do not touch the calibration panel directly. Please check the calibration panel if the calibration process
does not proceed normally. In case the calibration panel gets contaminated, contact the manufacturer
or your local service provider.
It is recommended to calibrate the device periodically
Go to Menu > Settings and configure the calibration period in Calibration Period (Days) option. The
default calibration period is 30 days.
3.1.1 Table Top Scanner Calibration
Turn the device on and make sure that it is discoverable by the program.
Click on the scanner icon in the bottom left corner to run the Calibration Wizard.
Prepare and place the calibration panel as shown on the picture.