SVERKER 750/760
Programma Electric AB
ZP-CD01E R05A
Auxiliary DC voltage output
egnaR
)CDV(
egatloV
)CDV(
tnerrucxaM
)CDAm(
031-0202003
04523
06553
08583
031-09004
022-031031523
051553
071583
022-081004
The auxiliary voltage output is stabilized and also
shortcircuit proof.
Moreover, it is equipped with two 400 V transient
voltage suppressors (transils). One is located
between the output contacts, and the other between
the black output contact and ground (earth)
The respective max currents can be used continu-
ously at high output voltages in either of the two
ranges. When the respective max currents are used
at low voltages in either range, a temperature
monitor will trip after a while, and the voltage will
be shut off. However, the voltage will be restored as
the temperature drops.
9.3 Inputs
Timer inputs
These inputs are equipped with 400 V transient
voltage suppressors (transils) at three locations:
• Between the input contacts.
• Between the black input contact and ground
(earth).
• Between the red input contact and ground
(earth).
Potential state
retemaraPniMepyTxaMtinU
revoegnahchgiH
CD|HIV|,CD,level
02CDV
revoegnahcwoL
CD|LIV|,CD,level
5CDV
,levelegatlovhgiH
CA|HIV|,CA
1
06smrCAV
,levelegatlovwoL
CA|LIV|,CA
3
tupnIsmrCAV
hgihtatnerruc
|HII|,CD/CA,level
15Am
.sgnidaeremittcerrocedivotpotdedeensiegatlovhgihsihT1
Contact state
retemaraPniMepyTxaMtinU
revoegnahchgiH
R,ecnatsiser
HI
3kΩ
revoegnahcwoL
R,ecnatsiser
LI
057
Ω
htiwtnerructuptuO
detiucric-trohs
tupni
731CDAm
ylppuslanretnI
Vegatlov
s
02CDV