6 Loading the steam sterilizer
23
Soft sterilization packaging
Soft sterilization packaging can be used in both sterilization containers and on trays. Please comply with
the following when using soft sterilization packaging e.g. MELAfol:
u Arrange soft sterilization packaging in a perpendicular position and at narrow intervals.
u Place transparent sterilization packages on their edge wherever possible and with the paper side
facing downwards.
u Do not place multiple soft sterilization packages flat on top of each other on a tray or in a container.
u If the seam seal tears during sterilization, this could be caused by the choice of undersized packaging.
Re-pack the instruments with larger packaging and perform sterilization again.
u Should the seam seal rip during sterilization, extend the sealing pulse on the film sealing device or
make a double seam.
Multiple wrapping
The steam sterilizer works with a fractionated vacuum procedure. This permits the use of multiple packag-
ing.
Mixed loads
Please observe the following when sterilizing mixed loads:
u Always place textiles at the top.
u Place the sterilization containers at the bottom.
u Place unwrapped instruments at the bottom.
u Place the heaviest loads at the bottom.
u Place transparent sterilization packaging and paper bags at the top - except in combination with
textiles. In this case, place them at the bottom.