14 Technical data
80
14 Technical data
Device type Vacuklav 41 B+ Vacuklav 43 B+
Device dimensions (H x W x D) 56.5 x 46 x 58 cm 56.5 x 46 x 69 cm
Empty weight 60 kg 69 kg
Operating weight 81 kg 91 kg
Sterilization chamber
Chamber diameter/depth Ø 25 cm | 35 cm Ø 25 cm | 45 cm
Chamber volume 18.4 l 23.8 l
Electrical connection
Electricity supply 220-240 V, 50/60 Hz, 3400 W
max. voltage range 207-253 V
Building fuses 16 A, RCD 30 mA
Length of the power cable 2 m
Overvoltage category (in
accordance with EN 61010-1)
Transient overvoltage up to the values of overvoltage category II
Degree of air pollution (in
accordance with EN 61010-1)
2
Surrounding conditions
Noise emission 64 dB(A)
Waste heat (with maximum load) 1.7 kWh
Ambient temperature 5-40 °C (ideal range 16-26 °C)
Degree of protection
(following IEC 60529)
IP20
Relative humidity max. 80 % at 31 °C, decreases in a linear fashion up to max. 50 %
relative humidity at 40 °C
Max. altitude 4000 m
Installation location Interior of a building
Feed water connection
Water quality EN 13060, Appendix C
Min. flow pressure 1.5 bar at 3 l/min
Min. static water pressure 2 bar
Max. static water pressure 10 bar
Max. water consumption 0.74 l 0.83 l
Wastewater connection
Max. through-flow volume 1.5 l 1.5 l
Max. water temperature 98 °C 98 °C