3 Performance specifications
9
whilst it opens. Pressure compensation will have been performed by the time that the door is completely
open, even following pressure differences.
Quantity and quality of the feed water
The quantity and quality of the feed water is automatically checked before every program start.
Overview of sterilization programs (type B)
The results in this table show which inspections were performed on the steam sterilizer. The marked field
shows compliance with all the applicable sections of the standard EN 13060.
Type tests Universal-
Program
Quick-
Program B
Quick-
Program S
Gentle-
Program
Prion-
Program
Program type in
accordance with
EN 13060
Type B Type B Type S Type B Type B
Dynamic pressure test of
the sterilization chamber
X X X X X
Air leakage X X X X X
Empty chamber test X X X X X
Solid load X X X X X
Porous partial load X -- -- X X
Porous full load X -- -- X X
Simple hollow body
(Hollow body B)
X X X X X
Product with narrow
lumen (Hollow body A)
X X -- X X
Single wrapping X X -- X X
Multiple wrapping X -- -- X X
Drying solid load X X X X X
Drying, porous load X -- -- X X
Sterilization temperature 134 °C 134 °C 134 °C 121 °C 134 °C
Sterilization pressure 2.1 bar 2.1 bar 2.1 bar 1.1 bar 2.1 bar
Sterilization time 5:30 min 5:30 min 3:30 min 20:30 min 20:30 min
X = Complies with all applicable sections of the standard EN 13060