2 Safety
6
2 Safety
When operating the device, comply with the following safety instructions as well as those
contained in subsequent chapters. Use the device only for the purpose specified in these
instructions. Failure to comply with the safety instructions can result in injury and/or damage to
the device.
Qualified personnel
n As with the preceding instrument decontamination, the sterilization of instruments and textiles using
this steam sterilizer may only be carried out by competent personnel.
Set-up, installation and commissioning
n Check the device for any damage suffered during transport after unpacking.
n The device should only be set-up, installed and commissioned by MELAG authorized technicians.
n The connections for electrical provision and water supply and discharge must be set-up by trained per-
sonnel.
n Using the optional electronic leak detector (water stop) minimizes the risk of water damage.
n The device is not suitable for operation in explosive atmospheres.
n Install and operate the device in a frost-free environment.
n The device is conceived for use outside the patient area. The device should be located a minimum of
1.5m radius away from the treatment area.
n The documentation media (computer, CF card reader etc.) must be placed in such a way that they
cannot come into contact with liquids.
n Observe all the information contained in the technical manual during commissioning.
Power cable and power plug
n Only the power cable included in the scope of delivery may be connected to the device.
n The power cable may not be replaced by a cable determined to be insufficient.
n Comply with all legal requirements and locally-specified connection conditions.
n Never operate the device if the plug or power cable are damaged.
n The power cable or plug should only be replaced by authorized technicians.
n Never damage or alter the power plug or cable.
n Never unplug by pulling on the power cable. Always take a grip on the plug.
n Ensure that the power cable does not become jammed in.
n Never lead the cable along a source of heat.
n Never fix the power cable with sharp objects.
Spring loaded safety valve
n The free movement of the valve rod must be guaranteed. For example, the spring safety valve must
not be taped or blocked. Install the device in such a way that the proper functioning of the spring load-
ed safety valve is guaranteed.
Reprocessing and sterilization
n Follow the manufacturer’s instructions of your textile articles and instruments regarding their decon-
tamination and sterilization.
n Comply with the relevant standards and directives applicable to the decontamination and sterilization
of textiles and instruments (in Germany e.g. from the RKI and DGSV).
n Only ever use packaging material and systems which have been cleared by their manufacturer for
steam sterilization.