EasyManua.ls Logo

MELAG Vacuklav 41 B+ Evolution - Sterilization; Important Information Regarding Routine Operation; Manufacturers Recommendation for Routine Operation

MELAG Vacuklav 41 B+ Evolution
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Sterilization
24
7 Sterilization
Important information regarding routine operation
Please comply with the recommendations issued by the Robert Koch Institute (RKI) and the information
contained in DIN58946-7.
See also “Routine Checks”.
Video tutorial
Manufacturer's recommendation for the routine operation of type B steam sterilizers
1)
When is it necessary to
make checks?
How should the checks be made?
Once per working day Visual check of the door seal and the door lock for damage.
Check the operating media (electricity, feed water and water connection
if necessary).
Check the documentation media (printer paper / computer / network).
We recommend performing the steam penetration test with MELAcontrol/
MELAcontrol PRO in the Universal-Program (test system in accordance
with DINEN 867-5).
Once a week Vacuum test
Tip: In the mornings before starting work - the steam sterilizer must
be cold and dry.
Batch-related tests With “Critical B” instruments:
MELAcontrol/MELAcontrolPRO must be used as batch control with
every sterilization cycle.
With “Critical A” instruments:
The process indicator (type 5 in accordance with ENISO11140) must
be used as batch control with every sterilization cycle.
With “Critical A + B” instruments:
MELAcontrol/MELAcontrolPRO must be used as batch control with
every sterilization cycle.
This simplifies the working procedure and increases security. You can
omit the daily steam penetration test with MELAcontrol/MELAcontrol PRO
(see above). The use of another test system in accordance with
DINEN867-5 is possible. The number of the available test systems
means that MELAG is not able to provide technical support when using a
different system.
PLEASE NOTE
The results of the tests must be documented.
n The indicator test strips used need not be stored.
1)
in accordance with the current recommendations from the Robert Koch Institute

Table of Contents

Related product manuals