7 Sterilization
30
2. Confirm the security query with YES.
3. After a short time, you can open the door by pressing
the symbol. The display shows a warning;
the log records the sterilization asNOT
successful.
Program abort after the start of drying
Should you abort a program after drying has started, the sterilization is having been completed successful-
ly. The steam sterilizer issues a malfunction message. You then need to expect insufficient drying, espe-
cially in the case of wrapped sterilized equipment and a full load. Sterile storage requires sufficient drying.
To ensure this, please allow programs with wrapped sterilized equipment to continue to the end of the dry-
ing phase as far as is possible. Unwrapped instruments sterilized in a Quick-Program dry after being re-
moved from their own warmth.
Proceed as follows to abort the program during drying:
1. Press STOP on the action bar.