11 Maintenance
64
2. Insert the new door seal into the groove at four points that are evenly distributed over the round blank.
3. Press the seal into the groove in each of the four quadrants. Ensure even distribution.
Housing parts
Where necessary, clean the housing parts with a neutral fluid cleaner or spirit.
Internal storage tank
Should you use the internal storage tank for the feed water supply, perform regular checks and cleaning as
follows:
Interval
Upon every refill Check the storage tank for soiling. If necessary, clean the storage tank with
fresh feed water before filling.
Every 2 weeks Clean the left-hand side chamber of the storage tank (wastewater).
Video tutorial
See also "Filling and emptying stand-alone autoclaves".
CAUTION
Hot wastewater in the storage tank and at the water drain tap.
This could result in burns
n Empty the storage tank before cleaning.
n Do not touch the water drain tap and parts connected to it with unprotected hands.
Empty the chamber of the internal storage tank
1. Open the service hatch in the lower area of the steam
sterilizer.
2. Place a min. 5 l bucket in front of the steam
sterilizer.
3. Connect the drain hose onto one of the drain fittings
on the waste water side (pos. a) or the feed water
side (pos. b) until it snaps into place. The knob must
be vertical.