Putting into Operation
21 D30374 11/2023
5. Putting into Operation
5.1 Putting into Operation for the First Time
WARNING
Condensation in the electrical components may cause short circuits.
Due to temperature fluctuations during transport, condensation may form inside the unit.
After transporting or storing the unit in humid conditions, remove it from its packaging
and allow it to acclimatise for at least 24hours in normal ambient conditions.
Do not connect the unit to the power supply during this time.
NOTICE
When putting the unit into operation for the first time, do not leave it unattended until it
has reached a steady state.
Please observe the national regulations when connecting the unit.
Observe the connected loads and power ratings (see }3.6Nameplate and
}3.7Technical Data).
Be sure to establish a safe PE conductor connection.
i
The temperature limiter might have been triggered during transport.
Press the red button on the back of the appliance to reset the temperature limiter
before putting the appliance into operation
See also
2 Nameplate [}13]
2 Technical Data [}14]
5.2 Connecting the Unit to the Power Supply
Route the power cable so that
nobody can trip over it.
it cannot come into contact with any hot parts.
it is easily accessible at all times and the plug can be pulled out quickly in the event of
a fault or emergency, for example.
5.3 Switching on Unit
1. Switch on the appliance by pressing the main switch on the front of the appliance.
ð The starting process is shown by three animated white dots (see
}7.1Warning Message of the Monitoring Function).
If the dots have another colour, an error has occurred (see
}7Malfunctions, Warning and Error Messages).