Malfunctions, Warning and Error Messages
36 D30374 11/2023
Description Cause Action
The appliance does not heat any more Mechanical temperature limiter (TB) has
switched off the heating permanently.
Allow the appliance to cool down.
Reset temperature limiter (TB). To do
this, press the red button on the rear
right of the appliance until you hear a
click.
If the alarm persists: Notify customer
service
The appliance does not cool any more or
cools more slowly
The fins on the cooling unit are dirty Clean the fins with a brush or hoover
7.2 Malfunctions, Operating Problems and Unit Errors
Error description Cause of errors Rectifying errors
Displays are dark External power supply was interrupted.
Check the power supply
Miniature fuse, appliance fuse or power
module faulty.
Notify customer service
Displays do not activate Appliance locked by USER ID.
Unlock with USER ID
Appliance is in programme, timer or remote
control mode (“Write” or “Write + Alarm”
mode).
Wait for the programme or timer to end
or switch off the remote control
Appearance of displays suddenly changes Appliance is in the “wrong” mode.
Press the MENU key to switch to the
operating or menu mode
Error message T:E-3 in the temperature
display
Temperature working sensor faulty.
Monitoring sensor performs the measuring
function.
The appliance can continue to be
operated for a short time
Notify customer service as soon as
possible
Error message AI E-3 in the temperature
display
Temperature monitoring sensor faulty.
Working sensor performs the measuring
function.
The appliance can continue to be
operated for a short time
Notify customer service as soon as
possible