Menu Mode
45 D30374 11/2023
4. Save the setting by pressing the confirmation key.
ð The corresponding calibration correction value is automatically highlighted.
5. Set the calibration correction value to 0.0 K.
6. Save the setting by pressing the confirmation key.
7. Position the sensor of a calibrated reference instrument centrally in the working
chamber of the appliance.
8. Close the door.
9. In manual mode, adjust the setpoint temperature.
10. Wait until the appliance reaches and displays the setpoint temperature.
■ The reference instrument will display the corresponding deviation.
11. In the SETUP, adjust the calibration correction value Cal2 to the deviation temperature
(actual value measured minus setpoint value).
12. Save the setting by pressing the confirmation key.
13. Compare the temperature measured by the reference measurement instrument with
the temperature displayed on the appliance.
ð After the calibration procedure, the temperature measured by the reference instrument
should now also be the setpoint temperature.
With Cal1, a further calibration temperature below Cal2, and with Cal3 a temperature
above, can be programmed in the same manner. The minimum difference between the Cal
values is 10K.
i
If all calibration correction values are set to 0.0 K, the factory calibration settings are
restored.
8.6 Programme
In the Programme display, programmes created using the AtmoCONTROL software can be
transferred to the appliance and saved on a USB storage medium. Here, you can also
select the programme you wish to run (see }6.5.3 Programme Mode) and delete
programmes.
1. Insert the USB storage medium on the right side of the ControlCOCKPIT.
ð You can now use one of the programmes saved on the USB storage medium.