44 D30374 11/2023
8.5 Calibrate
NOTICE
To guarantee problem-free closed-loop control, we recommend calibrating the appliance
once a year.
8.5.1 Temperature Calibration
The appliances are temperature calibrated and adjusted at the factory. If readjustment is
necessary for example due to the influence of the chamber load the appliance can be
calibrated for the specific customer using three possible calibration temperatures:
Cal1 Temperature calibration at low temperature
Cal2 Temperature calibration at medium temperature
Cal3 Temperature calibration at high temperature
i
For temperature calibration, you will need a calibrated reference instrument.
Example: Temperature deviation should be corrected
1. Press the activation key to the right of the CALIB display.
ð The display is enlarged and the temperature adjustment option is automatically
highlighted.
2. Press the confirmation key repeatedly, until the calibration temperature Cal2 is
highlighted.
3. With the turn control, set the calibration temperature Cal2 to the specified
temperature.