Operation and Control
33 D30374 11/2023
2. By turning the turn control, adjust the desired lower alarm limit.
i
The lower alarm limit cannot be higher than the upper alarm limit. If no undertemperature
protection is required, set the lowest temperature.
3. Press the confirmation key to confirm.
ð The max display (overtemperature protection) is activated.
4. By turning the turn control, adjust the desired upper alarm limit.
i
The monitoring temperature must be set sufficiently high above the maximum setpoint
temperature. We recommend 1 to 3 K.
5. Press the confirmation key to apply the setting for the upper alarm limit.
ð The setting of the automatic temperature monitor (ASF) is automatically activated
(auto).
6. With the turn control select ON (
) or OFF (
).
7. Press the confirmation key to confirm.
ð The ASF tolerance band setting is activated.
8. With the turn control, select the required tolerance band.
i
We recommend a tolerance band of 1 to 3 K.
9. Press the confirmation key to confirm.
ð Temperature monitoring is now active.
i
In menu mode you can set, whether an alarm should be accompanied by an acoustic
signal (see }8.7Acoustic Signals).