17
ZL6
3. SYSTEM
This section reviews the ZL6 data logger hardware features and configuration options. Use
this information to help optimize each environmental monitoring setup.
3.1 SPECIFICATIONS
MEASUREMENT SPECIFICATIONS
Sensor Logging Interval
ZL6 Basic
60 min (average or accumulation of 60, 1-min sensor
readings)
ZL6 and ZL6 Pro
5 min to 12 h (average or accumulation of 1-min sensor
reading)
Logger Reporting Interval
ZL6 Basic None
ZL6 Hourly with additional charges for more frequent reporting
ZL6 Pro Hourly with the ability to enable more frequent reporting
ZL6 Wi-Fi communication Upload after each measurement interval
Timekeeping
ZL6 Basic Synchronize with ZENTRA Utility or ZENTRA Utility Mobile
ZL6 and ZL6 Pro
Synchronize automatically and on-demand;
GPS, cellular or Wi-Fi communication, or ZENTRA Utility
software
COMMUNICATION SPECIFICATIONS
Computer Communication
Standard USB cable, USB A to micro-USB B
Cellular Communication (ZL6 and ZL6 Pro)
3G
Specifications UMTS 3G 5-band cellular module with 2G fallback