MI 3309 BT DeltaPAT Technical specifications
23
3.12. General data
Power supply voltage........................ 9 V
DC
(61.2 V NiMH or NiCd battery, type HR 6)
(size AA)
Operation.......................................... typical 8 h
Battery charging current.................... 250 mA (internally regulated)
Overvoltage category......................... CAT II / 300 V
Protection classification .................... double insulation
Pollution degree................................ 2
Protection degree case..................... IP 40
Protection degree test connectors .... IP 20
Display.............................................. 128 x 64 dots matrix display with backlight
Dimensions (w h d) ..................... 14 cm 8 cm 26 cm
Weight .............................................. 1.14 kg, without battery cells
Reference conditions:
Reference temperature range........... 10 C 30 C
Reference humidity range................. 40 %RH 70 %RH
Operation conditions:
Working temperature range .............. 0
C 40 C
Maximum relative humidity ............... 95 %RH (0
C 40 C), non-condensing
Storage conditions:
Temperature range ........................... -10
C +70 C
Maximum relative humidity ............... 90 %RH (-10
C +40 C)
80 %RH (40
C 60 C)
The error in operating conditions could be at most the error for reference conditions
(specified in the manual for each function) +1 % of measured value + 1 digit, unless
otherwise specified in the manual for particular function.
Memory..............................................1500 memory locations
Communication transfer speed:
RS232 interface ................................ 9600 bps, 1 start bit, 8 data bits, 1 stop bit
RS232 connector .............................. PS/2 connector, female
USB interface.................................... 115200 bps
USB connector.................................. type B
Bluetooth interface............................ 115200 bps
Protection pre-tests:
External voltage between LN and PE (DC and AC).
Excessive leakage between S/EB and PE (DC and AC).
L-N resistance is low or very low.
Connectivity (fuse) pre-test:
Appliance is not switched on or there is too high resistance between L and N.
Maximum resistance for connectivity pre-test..................................30 k

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the METREL DeltaPAT MI 3309 BT and is the answer not in the manual?

METREL DeltaPAT MI 3309 BT Specifications

General IconGeneral
BrandMETREL
ModelDeltaPAT MI 3309 BT
CategoryTest Equipment
LanguageEnglish

Related product manuals