137
8. Meter average .... xx UP/DOWN
Displays the number of measurements taken before the average value of build-in multimeter is displayed
(integration constant). UP, DOWN display buttons can be used to select a value in the range of 1 to 20. The
higher the value, the longer the calibrator needs to measure a value, but displayed value is more stable. If 20
is set, one measurement takes approximately 2.5 s.
9. Interface .... xx GPIB/RS232
10. GPIB address .... xx UP/DOWN
11. RS232 baud rate .... xx UP/DOWN
12. Handshake .... xx OFF/Xon-Xoff
Indicates the communication handshake. Display buttons can be used to select OFF or Xon/Xoff. Perfect
communication with the PC requires equal values set at the PC and the calibrator.
13. Keyb.beep .... xx ON/OFF
14. Keyb.volume .... xx UP/DOWN
15. Brightness .... xx UP/DOWN
16. Rotary change .... xx ON/OFF
This parameter controls the range of functions of the potentiometer. If ON is set, the potentiometer can move
the cursor both to the left and right (← and → symbols) and up and down (∧ and ∨ symbols). If OFF is set, the
potentiometer can move the cursor up and down only (∧ and ∨ symbols).
17. Switch polarity .... xx ON/OFF
This parameter controls the function of the built-in relay. If ON is set, the relay is closes before the test. If OFF
is set, the relay is opens before the test.
18. Switch activity .... xx PASS/FAIL
This parameter controls the function of the built-in relay. If PASS is set, the relay is active (changes its status)
if the test results in PASS status. If FAIL is set, the relay is active (changes its status) if the test results in FAIL
status.