152
25.3.2. SOURce subsystem
This subsystem allows to control the individual functions of CX 1651 calibrator.
Keyword Parameters
[SOURce]
: FUNCtion
[: SHAPe] (?) <CPD> { DC | SINusoid | PULPositive | PULSymmetrical |
PULNegative | RMPA | RMPB | TRIangle | LIMSinusoid |
PWMPositive | PWMSymmetrical | PWMNegative | .
SQUare }
: VOLTage
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: CURRent
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: RESistance
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: CAPacitance
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: POWEr
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: PHASe
: UNITS (?) <CPD> { DEG | COS }
[: ADJust] (?) <DNPD>
: VOLTage
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: CURRent
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: EARTh
: VOLTage (?) <CPD> { ON | OFF | 0 | 1 }
: CURRent (?) <CPD> { ON | OFF | 0 | 1 }
: AUXiliary (?) <CPD> { ON | OFF | 0 | 1 }
: ADAPter (?)
: FREQuency
[: CW ] (?) <DNPD>
: DUTY (?) <DNPD>
: VOLT (?) <DNPD>
: ATTE (?) <DNPD>
: TEMPerature
: UNITs (?) <CPD> { C | CEL | K }
: SCALe (?) <CPD> { TS68 | TS90 }