POWER FACTOR (PF)
range: -1.0 to +1.0
PF uncertainty can be calculated for any set value of output voltage, current and PF from following formula:
dPF = (1 – cos (ϕ+dϕ)/cos ϕ) * 100 [%]
where ϕ is set phase shift between voltage and current
dϕ is uncertainty of set phase shift form the above table
AC ENERGY uncertainty
It depends on set value of voltage, current, time and PF. The best uncertainty is 0.07% for apparent energy.
27.1.6. Frequency
summary range: 0.1 Hz to 20 MHz
frequency uncertainty: 0.005 %
output: BNC connector located on the front panel
modes: - PWM square wave output with calibrated duty cycle ratio, frequency and
amplitude
- HF square wave output with calibrated frequency and amplitude
Mode PWM
frequency range: 0.1 Hz to 100 kHz
voltage range: 1 mV to 10 V
duty cycle ratio range: 0.01 to 0.99
wave form: square, symmetrical– positive – negative
duty cycle ratio uncertainty: 0.05 %
Amplitude uncertainty
range % value + % range
Mode HF
frequency range: 0.1 Hz to 20 MHz
output impedance: 50 Ω
wave form: square symmetrical, duty cycle ratio 1:1
amplitude: 4 V
pk-pk
output amplitude range: 0, -10, -20, -30 dB +/- 1 dB
amplitude uncertainty: 10 %
rise/fall time: < 3 ns