Operation
64
Low fill levels
When the fill level is low in one of the containers, e.g., for process
chemicals or the salt container, you are prompted to refill the con-
tainer.
Acknowledge the messages with OK and refill the relevant contain-
ers. To do this, follow the instructions in the sections “Filling
process chemicals and dispensing” and “Water softener”.
Spray arm pressure and monitoring
The machine has a sensor for monitoring spray pressure e.g in order
to detect pressure fluctuations caused by improper loading of the
machine or foam in the water circulation system. Spray pressure
monitoring is activated for the “Cleaning” and “Final rinse” wash
blocks.
The spray pressure monitoring result is documented together with the
process documentation.
Spray arm speed can also be monitored, e.g. for detection of block-
ages caused by improper loading of the machine or foam in the water
circulation system. Spray arm speed monitoring can be activated or
deactivated via the programmable settings.
Miele Service can make further settings for spray pressure and spray
arm monitoring.