PRE INSTALLATION
Chapter 3 PRE INSTALLATION
3-1 INSTALLATION REQUIREMENTS
FEEDWATER REQUIREMENTS
Type of Feedwater Potable
Flowrate 30 LPH
Minimum Pressure 0.5 bar
Maximum Pressure 6 bar
Conductivity < 2000 μS/cm
Temperature 5 to 35 °C
pH 4 - 10
Fouling Index < 10
Iron < 0.1 ppm as CaCO
3
Aluminum < 0.05 ppm as CaCO
3
Manganese < 0.05 ppm as CaCO
3
Free Chlorine < 1 ppm
Langelier Saturation Index < + 0.2
TOC < 2000 ppb
FEEDWATER CONNECTION NEEDED
Feedwater Piping Connection 1/2 inch Male GAZ, NPTM or BSPM
REJECT FLOW REQUIREMENT
Drain Capacity 30 LPH
- 6 -