MAINTENANCE
6-5 HOW TO SANITIZE THE TANK
A tank sanitization is performed to eliminate bacteria growth in the tank. It is recommended to sanitize the
tank at least once a year.
T
HINGS TO KNOW BEFORE YOU SANITIZE THE TANK
All safety precautions must be followed when
handling Hydrogen Peroxide. Rubber gloves,
safety goggles and a lab coat must be worn to
avoid any skin and body contact.
!
HAZARD
You will need a minimum Tank Level display of at least 50% (A).
Locate the clear elbow fitting, the clear tubing and the syringe in the Sanitization Kit (B).
You will need 200 ml of 30% of Hydrogen Peroxide solution and 200 ml of purified water.
Millipore does not sell Hydrogen Peroxide but it is readily available through most Scientific Supply
Companies.
The Vent Filter will need to be replaced after the sanitization is completed.
The system will not be able to deliver RO water while the system is being sanitized.
ATTENTION
The total time needed to sanitize the tank is at
least 4.5 hours.