Bacteria | < 1 CFU/mL |
---|---|
Pyrogens (Endotoxins) | < 0.001 EU/mL |
Type | Water purification system |
Filtration Technology | Reverse osmosis, ultrafiltration |
Output Water Quality | Type I ultrapure water |
Resistivity | 18.2 MΩ·cm |
TOC | ≤ 5 ppb |
Flow Rate | Up to 2 L/min |
Feed Water Quality | Potable water |
RNases | Not detected |
DNases | Not detected |
Particle Size | 0.22 μm |
Dimensions | Varies by model |
Weight | Varies by model |
Power Supply | 100-240 V, 50/60 Hz |
Guide for installation, operation, and maintenance of Synergy/Synergy UV-R systems.
Guidance on operating the Synergy R System according to manual instructions and specifications.
Information on finding addresses, telephone/fax numbers, and other details via the Millipore SAS Internet Site.
Visual overview of the Synergy R system components: Display, Dispensing Button, and Keypad.
Details on water flowrate, resistivity, TOC, and micro-organisms for Synergy R product water.
Diagram illustrating key components of the Synergy R system with numbered labels.
Explanation of how the Synergy R Water Purification System functions with its components.
Detailed specifications for Synergy R system dimensions, weight, noise level, and electrical requirements.
Specifies feedwater requirements (type, flowrate, pressure, temp, conductivity, TOC) and environmental conditions.
Details the required feedwater piping connection type (1/2 inch Male GAZ, NPTM or BSPM).
Lists requirements for storage/operating temperature, relative humidity, altitude, installation category, and pollution degree.
Information on available accessories like mounting brackets and supports for the remote dispenser.
Instructions on opening the shipping box and using the accessories checklist for unpacking.
Step-by-step guide on connecting the feedwater tubing to the system and feedwater source.
Instructions for connecting the ultrapure water dispenser, including tubing and electrical connectors.
Procedure for connecting the power cord, powering on the system, and opening the feedwater supply valve.
Detailed steps for installing the SynergyPak cartridge, including safety precautions.
Instructions for installing the Tygon tubing and barbed fitting onto the POU dispenser.
Steps to purge trapped air from the SynergyPak cartridge using the dispensing button.
Procedure for hydrating the system, involving leaving it overnight in PRE OPERATE mode.
Guide for installing and rinsing the Final Filter, ensuring proper sealing and air venting.
Step-by-step instructions for calibrating the product flowrate using a graduated cylinder.
Guide on selecting and displaying resistivity or conductivity units, with or without temperature compensation.
Instructions on setting the resistivity setpoint to alert users when product resistivity is low.
Explanation of the display elements showing operating modes, parameters, and maintenance/alarm messages.
Describes two methods of dispensing water using the Dispensing Button: press and hold, or press and release.
Guide on dispensing a precise volume of product water using the F01 menu option.
Details the different operating modes: Standby and Pre Operate, including their functions and display indicators.
Describes the Dispensing mode, when it is active, and what is displayed.
Explains the Auto-Dispensing mode used for dispensing pre-set volumes, and what the display shows.
Instructions on viewing resistivity and temperature values displayed during Pre Operate mode.
Guide on manually recirculating water to enhance quality before dispensing.
Explains how to interpret Pack Alarm and UV Lamp Alarm messages on the system display.
A schedule outlining recommended maintenance tasks, their triggers, and related section references.
Step-by-step instructions for safely removing and replacing the SynergyPak cartridge.
Detailed procedure for removing and installing a new Final Filter, with attention to sealing.
Instructions for cleaning the screen filter to prevent blockages and maintain feedwater flow.
Step-by-step guide for replacing the UV Lamp, including safety precautions.
Instructions on how to view the remaining days on the UV Lamp Timer and reset it.
Detailed steps for resetting the UV Lamp Timer after replacement.
Troubleshooting steps for a blank display screen, checking power and fuses.
Steps to resolve low or no water flow, checking filters and air locks.
Solutions for inaccurate auto-dispensing, focusing on flowrate calibration and filter issues.
Troubleshooting for missing resistivity values, checking measurement conditions and cartridge status.
Actions to take when the SynergyPak cartridge lifetime is exhausted or resistivity is low.
Steps to address incorrect or missing SynergyPak cartridge installation.
Steps to take if the system has been in standby mode or resistivity is below setpoint.
Actions for low product resistivity, including recirculation and cartridge replacement.
What to do when the UV Timer is exhausted, including lamp replacement and timer reset.
Troubleshooting for improperly installed UV lamps, suggesting power off and reinstallation.
Provides catalogue numbers for Synergy R systems based on voltage configurations.
Lists consumable items for Synergy R systems with their corresponding catalogue numbers.
Lists available accessories for the Synergy R system and their catalogue numbers.
Reference to page 26 for checking water quality.
Reference to page 2 for contacting Millipore SAS.
Reference to page 23 for delivering a fixed volume of ultrapure water.
Reference to page 23 for delivering ultrapure water.
Reference to page 26 for improving water quality via recirculation.
Reference to page 33 for replacing the Millipak end filter.
Reference to page 30 for replacing the SynergyPak.
Reference to page 35 for replacing the UV lamp.
Reference to page 2 for online e-learning resources.