EasyManua.ls Logo

Millipore RIOS 3 - Page 40

Millipore RIOS 3
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAINTENANCE
- 33 -
Open the front cover to go into STANDBY mode (I).
Let the system stand for 1 hour for effective bacteria elimination.
(After 1 hour) Check that the Tank Outlet Tubing is secured into the drain. Open the Tank Outlet
Valve to drain all the water from the tank (J).
Close the Tank Outlet Valve (K).
Close the front cover (L). The system will now go into FILLING TANK mode. Let the tank level fill up
to the 100% level.
I
J
K
L
Repeat steps I, J, K and L:
Open the front cover to go into STANDBY mode (I).
Check that the Tank Outlet Tubing is secured into the drain. Open the Tank Outlet Valve to drain all
the water from the tank (J).
Close the Tank Outlet Valve (K).
Close the front cover (L). The system will now go into FILLING TANK mode.
The tank is now sanitized.

Table of Contents

Related product manuals