Questions and Answers

  • D
    Denise EverettSep 23, 2025
    What to do if Milltronics Measuring Instruments doesn't respond to programmer?
    • C
      Christopher TaylorSep 23, 2025
      If your Milltronics Measuring Instruments are not responding to the programmer, the infrared interface may be obstructed, or the programmer may be defective. Clean the enclosure "docking bay" and the programmer magnet slot.
  • M
    Michelle AndersonAug 23, 2025
    What to do if Milltronics Control Unit reading is erratic?
    • J
      jgrayAug 25, 2025
      If the reading is erratic with little or no relation to the material level, the true echo may be too weak or the wrong echo is being processed. Try to relocate and / or reaim transducer at material level or object.
  • M
    middletonstacieAug 25, 2025
    What to do if Milltronics Control Unit displays 'EEEE'?
    • R
      Rachel ReillyAug 26, 2025
      If the Milltronics Control Unit displays 'EEEE', it means the reading is too high. To resolve this, select larger units (P005) or lower the Convert Reading value (P061).
  • E
    Erica CastroAug 27, 2025
    What to do if my Milltronics Control Unit always reports the material level 'off'?
    • T
      todd62Aug 27, 2025
      If the material level reported is always 'off' by the same amount, the empty (zero) reference is incorrect. You should check Empty (P006), Reading Offset (P063), Offset Calibration (P650), & Offset Correction (P652).
  • J
    Jack RivasAug 27, 2025
    What to do if Milltronics Control Unit reading is fixed?
    • R
      Robert HaleAug 27, 2025
      If the reading is fixed, regardless of the actual material level, the transducer beam may be obstructed, the standpipe is too narrow, or the transducer mounting is resonant. Relocate and / or reaim transducer at material level or object.
  • M
    Megan EdwardsSep 15, 2025
    How to troubleshoot a blank display on Milltronics Measuring Instruments?
    • J
      Jose SpenceSep 15, 2025
      If your Milltronics Measuring Instruments display is blank, the transmit neon is not flashing, and the transducer is not pulsing, it's likely due to a lack of power. Check the power supply, wiring, and voltage selection switch.
  • R
    Russell BoydSep 12, 2025
    Why does the reading fluctuate on Milltronics Measuring Instruments?
    • J
      Jill YangSep 12, 2025
      If the reading on your Milltronics Measuring Instruments fluctuates while the material level is still (or vice versa), it indicates incorrect measurement stabilization. Adjust the Measurement Response (P003) accordingly.
  • J
    John SmithSep 14, 2025
    Why is the reading erratic on Milltronics Measuring Instruments?
    • J
      James WardSep 14, 2025
      If the reading on your Milltronics Measuring Instruments is erratic and shows little to no relation to the material level, it suggests that the true echo is too weak or the wrong echo is being processed. Try relocating and/or re-aiming the transducer at the material level or object.
  • T
    Teresa TannerSep 5, 2025
    What does 'Short' and 'tb:(#)' mean on my Milltronics Control Unit?
    • T
      Tracey GravesSep 5, 2025
      If the Milltronics Control Unit displays 'Short' and 'tb:(#)', it indicates a short circuited transducer cable or a defective transducer. Repair or replace as necessary.
  • A
    April AlvaradoSep 2, 2025
    What does 'Open' and 'tb:(#)' mean on my Milltronics Control Unit?
    • R
      rachel76Sep 2, 2025
      If the Milltronics Control Unit displays 'Open' and 'tb:(#)', it indicates an open circuited transducer cable, or a defective transducer. Repair or replace as necessary.

Summary

Installation

Location, Mounting, and Cable Entry

Covers ideal mounting locations, physical mounting, and cable entry requirements.

Programming

Entering Program Mode and Value Alteration

Steps to enter programming and change parameter values.

Quick Start Parameters

Initial Configuration Settings

Sets up operation mode, material, response, transducer, units, empty, and span.

Operation

Performance Evaluation and Testing

Steps to verify system performance and complete setup.

Application Parameters

Reading and Failsafe Configuration

Adjusts reading display and configures failsafe behavior.

Relay Configuration

Sets up relay alarms and setpoints.

mA Output Configuration

Configures mA outputs and their limits.

Enhancement Parameters

Troubleshooting Guide

Symptoms, Causes, and Actions

A table listing common issues and solutions.

Measurement Difficulties and LOE Display

Troubleshooting common measurement problems and the LOE indicator.

Fixed and Wrong Readings

Diagnosing and resolving fixed or incorrect readings.

Milltronics airanger XPL plus Specifications

General IconGeneral
ManufacturerSiemens
TypeUltrasonic Level Transmitter
Frequency50 kHz
Process PressureUp to 3 bar
Accuracy± 0.25% of range or 6 mm (0.25 inch), whichever is greater
Process Temperature-40 to +85 °C (-40 to +185 °F)
CommunicationHART
Beam Divergence10°
Output4-20 mA
Weight1.5 kg
Operating Temperature-40 to +80 °C (-40 to +176 °F)