13 - 4 Defibrillator/Monitor Operator’s Manual
WARNING
• Continuous non-invasive blood pressure measurements may cause purpura, ischemia and
neuropathy in the limb with the cuff. Inspect the application site regularly to ensure skin quality and
inspect the extremity of the cuffed limb for normal color, warmth and sensitivity. If any abnormity
occurs, move the cuff to another site or stop the blood pressure measurements immediately.
13.6.6 Starting a STAT Measurement
1. Select the NIBP parameter area to enter the [NIBP Setup] menu.
2. Select [NIBP STAT]. The STAT mode initiates a 5-minute continuous, automatic NIBP measurement.
13.7 Setting Initial Cuff Inflation Pressure
You can set the initial cuff inflation pressure manually.
1. Select the NIBP parameter area to enter the [NIBP Setup] menu.
2. Select [Initial Pressure] and then select the appropriate setting. The cuff is inflated accordingly at the next
NIBP measurement.
NOTE
• For known hypertensive patients, you need to set initial cuff pressure to a higher value to reduce the
measurement time.
• Setting initial cuff inflation pressure is disabled during NIBP measurement.
• The initial cuff inflation pressure is restored to the default setting if NIBP module has been reset or
patient category has been changed.
13.8 Setting Pressure Unit
You can set [Press. Unit] to change NIBP unit only by selecting [NIBP Setup] from the Configuration Main menu.
13.9 NIBP Troubleshooting
For details, refer to EAlarm Messages.

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)

Related product manuals