Defibrillator/Monitor Operator’s Manual 12 - 3
Interfering substances:
Intravascular dyes (such as indocyanine green, methylene blue, indigo carmine, etc.)
Dyes in the measure site, such as nail polish.
Environmental conditions:
Excessive ambient light
Electrosurgery equipment. The pulse oximeter may be used during electrocautery, but this may affect
the accuracy or availability of the parameters and measurements.
Defibrillation (may cause inaccurate reading for a short amount of time)
Excessive patient/sensor motion
Electromagnetic field
Arterial catheters and intra-aortic balloon
Others
Inappropriate positioning of the SpO
2
sensor, or use of incorrect SpO
2
sensor
Cuff or arterial blood pressure measurement device on the same limb as the SpO
2
sensor.
12.5 SpO2 Display
NOTE
PI is only available for Masimo SpO
2
.
(1) Pleth waveform (Pleth): visual indication of patient’s pulse. The waveform is not normalized.
(2) SpO
2
unit
(3) SpO
2
alarm limits
(4) Oxygen saturation of arterial blood (SpO
2
): percentage of oxygenated hemoglobin in relation to
the sum of oxyhemoglobin and deoxyhemoglobin.
(5) Perfusion indicator: the pulsatile portion of the measured signal caused by arterial pulsation.
(6) Perfusion index (PI): gives the numerical value for the pulsatile portion of the measured signal
caused by arterial pulsation. PI is an indicator of the pulsatile strength. You can also use it to
assess the SpO
2
signal strength.
If the “SpO2 Low Perf alarm persists, reposition the SpO
2
sensor or find a better site.
If low perfusion persists, choose another method to measure oxygen saturation if possible.
(1)
(2)
(3)
(6)
(4)
(5)

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)

Related product manuals