21 - 10 Defibrillator/Monitor Operator’s Manual
21.6.11 CO
2
Setup Menu
Dia High Adu (Low+5) to 250 90 Step: 1 mmHg
Ped (Low+5) to 200 70
Neo (Low+5) to 115 60
Dia Low Adu 10 to (High-5) 50
Ped 40
Neo 20
Menu Item Options/Range Default Description
Menu Item Options/Range Default Description
Alarm On, Off On /
Alm Lev High, Med Med /
Press. Unit mmHg, kPa, % mmHg Sets the CO
2
measurement unit.
Alarm limits are refreshed in real
time when pressure unit is
changed.
Sweep 6.25 mm/s, 12.5 mm/s, 25
mm/s
6.25 mm/s Sets the CO
2
waveform speed.
Scale mmHg 15, 20, 25, 40,
50, 60, 80
50 Different options are available
to match the selected pressure
unit.
kPa 2, 2.5, 3.5, 5, 7,
8, 10
7
% 2, 2.5, 3.5, 5, 7,
8, 10
7
EtCO
2
High Adu (Low+2) to 99 50 Step: 1 mmHg
The range is different according
to the selected pressure unit.
The default pressure unit is
mmHg.
Ped 50
Neo 45
EtCO
2
Low Adu 1 to (High-2) 25
Ped 25
Neo 30
FiCO
2
High Adu 1 to 99 4
Ped 4
Neo 4
awRR High Adu (Low+2) to 100 30 Step: 1 rpm
Ped 30
Neo (Low+2) to 150 100
awRR Low Adu 6 to (High-2) 8
Ped 8
Neo 30
Apnea Time 10 s, 15s, 20s, 25s, 30s, 35s,
40s
20s The equipment gives an apnea
alarm if the patient has stopped
breathing for longer than the
preset apnea time.

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)