F - 2 Defibrillator/Monitor Operator’s Manual
F.3 Device Labeling
Check the labels provided by the manufacturer or the healthcare facilities are present and legible.
â–  Main unit label
â–  Integrated warning labels
F.4 Protective Earth Resistance
1. Plug the probes of the analyzer into the device’s protective earth terminal and protective earth terminal of
the AC power cord.
2. Test the earth resistance with a current of 25 A.
3. Verify the resistance is less than limits.
LIMITS
For all countries, R = 0.2  Maximum
F.5 Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage tests.
The following outlet conditions apply when performing the Earth Leakage test:
â–  normal polarity (Normal Condition)
â–  reverse polarity (Normal Condition)
â–  normal polarity with open neutral (Single Fault Condition)
â–  reverse polarity with open neutral (Single Fault Condition)
LIMITS
For UL60601-1,
◆ 300 A in Normal Condition
◆ 1000 A in Single Fault Condition
For IEC60601-1,
◆ 500 A in Normal Condition
◆ 1000 A in Single Fault Condition
F.6 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All measurements have
a true RMS only
The following outlet conditions apply when performing the Patient Leakage Current test.
â–  normal polarity (Normal Condition)
â–  reverse polarity (Normal Condition)
â–  normal polarity with open neutral (Single Fault Condition)
â–  reverse polarity with open neutral (Single Fault Condition)
â–  normal polarity with open earth (Single Fault Condition)
â–  reverse polarity with open earth (Single Fault Condition)
LIMITS
For CF applied parts
◆ 10A in Normal Condition
◆ 50A in Single Fault Condition

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)