Defibrillator/Monitor Operator’s Manual D - 1
D Defibrillator Shift Checklist
Equipment Name: Serial Number: Department:
Item Pass/
Fail
Corrective Actions/Remarks
1. Equipment Appearance
â—† Clean, no foreign substance, no crack
2. Cables/Connectors
â—† Cables not frayed, connectors and pins not broken or
loose
â—† Connectors engage securely
3. Batteries
â—† Battery installed with at least 60% of battery capacity
â—† Fully charged spare battery available
4. Necessary Accessories (Electrode pads, electrodes,
paddles or recorder paper)
â—† Present and sufficient
â—† Inspected to be used normally
5. Auto Test
â—† Status indicator off
6. Shock Test
â–  If the external paddles are used:
a. Connect the equipment with AC power supply, and
the AC power indicator is illuminated.
b. Connect the paddles cable to the equipment, and
place the external paddles in the paddle tray.
c. Press the Charge button on the external paddles, and
charge the equipment to 50J.
d. Press the Shock button on the external paddles.
e. The system prompts that the shock is delivered
normally.
6. Shock Test
â–  If the electrode pads are used:
a. Connect the equipment with AC power supply, and
the AC power indicator is illuminated.
b. Connect the pads cable to the equipment.
c. Perform the Energy Delivery test item of the user test
with the test load connected.
d. The system prompts that the Energy Delivery test is
passed.
7. Monthly Check on Expiration Date
â—† The electrode pads are not expired.
Checked by:
Date:

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)

Related product manuals