21 - 12 Defibrillator/Monitor Operator’s Manual
21.6.14 Mark Event Setup Menu
21.6.15 Record Setup Menu
Menu Item Options/Range Default Description
Event A Generic Generic Unchangeable
Event B Adrenalin, Lidocaine,
Atropine, Nitroglycerin,
Morphine, Intubation, IV
Access, Adenosine,
Amiodarone, Vasopressin,
Isoprenaline, Dopamine,
Aspirin, Oxygen, CPR
Adrenalin Event names that have been
selected by previous events will
not be included in the options
of later events.
Event C Lidocaine
Event D Atropine
Event E Nitroglycerin
Event F Morphine
Custom Event 1 Options: entering event
name using the keyboard
included in the [Mark Event
Setup] menu.
Range: 1 to 20 characters
/ After being defined, custom
event is refreshed in real-time in
the Mark Event list.
Custom Event 2
Menu Item Options/Range Default Description
Wave Length 8s, 16s, 32s, STAT 8s Sets the recording time of
recorded waveform.
Paper Speed 6.25 mm/s, 12.5 mm/s,
25mm/s, 50mm/s
25mm/s Sets the recording speed.
Energy Delivered Output On, Off Off Sets whether the energy level of
a delivered shock is displayed in
the event recording.
Gridlines On, Off On Sets whether the recording
waveforms are displayed with
gridlines.
Auto Record Charge Event On, Off Off Sets whether the event is
automatically recorder when
the related event is triggered.
Shock Event On
Marked Event Off
Auto Test
Report
On, Off, Only if Failed Off Sets whether the auto test
report is automatically recorded
when the auto test is
completed.
•[On]: the auto test report is
automatically recorded
when the auto test is
completed.
•[Off]: the auto test report is
not recorded when the auto
test is completed.
•[Only if Failed]: the auto test
report is automatically
recorded when the auto test
is completed and the auto
test fails.

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)

Related product manuals