EasyManuals Logo

Mindray BeneHeart D3 User Manual

Mindray BeneHeart D3
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
14 - 6 Defibrillator/Monitor Operator’s Manual
14.7 Removing the Exhaust Gases from the System
WARNING
• When taking the CO
2
measurement on patients who are receiving or have recently received
anesthetics, connect the outlet to a scavenging system, or to the anesthesia machine/ventilator, to
avoid exposing medical staff to anesthetics.
To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet connector of the
module.
14.8 CO
2
Calibration
CO
2
calibration should be performed by Mindray-qualified service personnel only once a year or when the
readings go far beyond the range. For details, refer to the relevant service manual.
CAUTION
• Connect the gas outlet to the scavenging system when calibrating CO
2
.
14.9 CO
2
Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or
accessories, check the table below before requesting for services. If the problem persists, contact your service
personnel.
NOTE
• For the physiological and technical alarm messages, refer to EAlarm Messages.
Problem Corrective Actions
EtCO
2
measurements too low 1. Ventilate the room if the environmental CO
2
concentration is too high.
2. Check the sample line and connectors for leakage.
3. Check the patient status.

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)

Related product manuals