24 - 4 Defibrillator/Monitor Operator’s Manual
24.3.2.2 Viewing the User Test Summaries
The results of the user test are automatically saved as summaries. The equipment can store up to 300 historical
test summaries which are listed in the sequence of time, with the latest on the top.
1. Press the Main Menu button on the front panel, and then select [User Test >>] [Ye s ].
2. Select [History].
3. Select the desired item to check the result details.
NOTE
If the routine test item (external discharge) of the auto test, or the energy delivery test item
(external discharge) of the user test is passed, the delivered energy accuracy is displayed, but it is for
your reference only.
24.3.2.3 Transmitting the User Test Summaries
You can also send the user test summaries to the CMS.
1. Press the Main Menu button on the front panel, and then select [User Test >>] [Yes].
2. Enter the [Transmit] menu by either of the following ways.
Select [Start], and select [Transmit] at the completion of a user test.
Select [History], select a desired item, and select [Transmit].
3. Set [Select Site].
A preset site: [Site Address] is inputted automatically. For details on configuring a preset site, refer to
20.2.2Storing Preset Sites
.
[Custom]: [Site Address] needs to be inputted manually.
4. Select [Transmit].
The message “Transmission Complete! displays if user test summaries are successfully transmitted.

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)

Related product manuals