A - 16 Defibrillator/Monitor Operator’s Manual
• When the equipment and related products have differing environmental specifications, the
effective range for the combined products is that range which is common to the specifications for all
products.
NOTE
• The environmental specification of unspecified modules are the same as those of the main unit.
Main Unit
Item Temperature Relative humidity Barometric
Operating condition 0ºC to 45ºC
(at least 60 minutes of working
time when the temperature
reduces from room temperature
to -20ºC)
10% to 95%,
non-condensing
-381m to 4575m
(57kPa to 106.2kPa)
Storage condition -30ºC to 70ºC
CO
2
module
Item Temperature Relative humidity Barometric
Operating condition 5ºC to 40ºC 10% to 95%,
non-condensing
430mmHg to 790 mmHg
(57.3kPa to 105.3kPa)
Storage condition -20ºC to 60ºC
Shock
Complies with requirements of 21.102, ISO9919:
Peak acceleration: 1000m/s
2
(102g)
Duration: 6ms
Pulse shape: half-sine
Number of shocks: 3 shocks per direction per axis (18 total)
Vibration
Complies with requirements of 21.102, ISO9919.
Bump
Complies with the requirements of 6.3.4.2, EN1789.
Peak acceleration: 15g
Duration: 6ms
Number of impacts: 1000
Impact direction: vertical impacts are applied when the equipment under test is placed at normal operating position.
Free fall
Complies with the requirements of 6.3.4.3, EN1789.
Drop height: 0.75 m
Number of drops: once for each of the six surfaces

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)