System Diagnosis and Support 11-11
11.3.7 Other Alarms
on; if not, change a new gel warmer.
The status indicator
blinks in orange in a
frequency of 0.5 Hz.
The gel warmer is not able to
heat due to the damage of
internal components.
Unplug and reconnect the power
cord to observe if the orange light
blinks in frequency of 0.5Hz; if yes,
change a new gel warmer.
The status indicator
blinks in orange in a
frequency of 2Hz.
Overheating protection due to
damage of internal components.
Disconnect the power cord and wait
about 10 minutes for the cooling of
the gel warmer, reconnect the power
cord, start to heat, wait about 10
minutes before the gel warmer
becoming warm, and observe if the
orange light blinks in frequency of 2
Hz; if yes, change a new gel warmer.
The status indicator
lights in green, but the
gel warmer is not warm
at all
Failure in the internal electrical
connection.
Disconnect and reconnect the power
cord, start to heat, wait about 10
minutes and touch the gel warmer to
see if it is warm; if not, replace a new
gel warmer.
"SystemConfiguration.ini"
can’t be opened, please
check the HDD data!
ADT7462 initialization is
abnormal
XXX ADT7462 initialization error:
XXX ADT7462, details are:
PC Board A,
PC Board B,
PC Board A & B,
Front-End Power Main Board C,
PC Board A & Front-End Power Main Board C,
PC Board B & Front-End Power Main Board C,
PC Board A & B & Front -End Power Main Board C
Reinstall the
system
software/check if
the drive is
normal/check if
the digital board is
normal/check if
the front-end
power main board
is normal.