Care & Maintenance 12-1
12 Care & Maintenance
12.1 Overview
The following procedures are recommended.
12.1.1 Tools, Measurement Devices and Consumables
Table 12-1 Tools and Measurement Devices
Tool/Measurement Devices
Qty.
Remarks
Resin or plastic container
1 pcs
Can accommodate two probes
Soft brush
1 pcs
About a toothbrush size
Small plastic basin
1 pcs
Used to fill the soapy water
Safety test analyzer
1 pcs
Refer to appendix A
Inner hexagon wrench
2 pcs
Inner hexagon wrench 8 and 6
Table 12-2 Consumable List
12.1.2 Care and Maintenance Items
Table 12-3 Maintenance Items
NO.
Maintain content
Frequency
Method
1.
Clean dust-proof covers
Once a month
Referring to12.2.1
2.
Clean monitor and touch screen
Once a month
Same as the above
Consumable
Qty.
Remarks
Aluminum foil
About 1 meter
Physiological saline
About 1000ml
Filling a half container Immerging the
whole probe (referring to appendix A).
(concentration 0.85 0.95%)
Mild soapy water
About 400ml
Dry soft cloth/cotton cloth
About 5 pcs

Table of Contents